Remote control commands
R&S
®
RTE
1342User Manual 1326.1032.02 ─ 20
MTESt:WFMLupdate <MaskTestName>
Creates the upper and lower mask limit from the envelope of the selected reference
waveform. If the reference waveform was not defined before, it is created automatically
from the mask test source waveform which is set wtih MTESt:SOURce.
Setting parameters:
<MaskTestName> String containing the name of the mask test
Usage: Setting only
Asynchronous command
MTESt:WFMRescale:XWIDth <MaskTestName>,<HorizontalWidth>
MTESt:WFMRescale:XWIDth? <MaskTestName>
Sets the width of the mask in horizontal direction. The specified factor in divisions is
added to the positive x-values and subtracted from the negative x-values of the mask
limits in relation to the source waveform of the mask.
Parameters:
<HorizontalWidth> Range: 0 to 1000
Increment: 0.01
*RST: 0
Default unit: div
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
Usage: Asynchronous command
MTESt:WFMRescale:YWIDth <MaskTestName>,<VerticalWidth>
MTESt:WFMRescale:YWIDth? <MaskTestName>
Sets the width of the waveform mask in vertical direction. The specified factor in divi-
sions is added to the y-values of the upper mask limit and subtracted from the y-values
of the lower mask limit. Thus, the upper half of the mask is pulled upwards, the lower
half is pulled down.
Parameters:
<VerticalWidth> Vertical mask width in divisions
Range: 0 to 1000
Increment: 0.01
*RST: 0
Default unit: div
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
Usage: Asynchronous command
Mask testing