Remote control commands
R&S
®
RTE
1467User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<PrimaryBitRate> Range: 0 to 100000000000
Increment: 1
*RST: 0
Default unit: bps
Usage: Query only
BUS<m>:SPI:FRAMe<n>:STATus?
Returns the overall state of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameState> OK | VOID | INCFirst | INCLast | INSufficient
OK: the frame is valid.
VOID: the frame is empty.
INCFirst: INComplete First word. The first word does not have
the expected word length.
INCLast: INComplete Last word. The last word does not have
the expected word length.
INSufficient: The frame is not completely contained in the acqui-
sition. The acquired part of the frame is valid.
*RST: OK
Usage: Query only
BUS<m>:SPI:FRAMe<n>:STARt?
Returns the start time of the specified frame.