Remote control commands
R&S
®
RTE
1495User Manual 1326.1032.02 ─ 20
If the trigger type is set to ID (identifier), you can set the values
ANY | DATA | REMote.
*RST: ANY
TRIGger<m>:CAN:ITYPe <IdentifierType>
Selects the format of data and remote frames.
Remote frames are not available in the CAN FD protocol.
Parameters:
<IdentifierType> ANY | B11 | B29
B11
11 bit identifier (standard format). The instrument triggers on the
sample point of the IDE bit.
B29
29 bit identifier (extended format). The instrument triggers on the
sample point of the RTR bit.
ANY
The ID type and ID pattern are not relevant for the trigger condi-
tion.
*RST: ANY
TRIGger<m>:CAN:ICONdition <IdOperator>
Sets the operator to set a specific identifier or an identifier range.
Parameters:
<IdOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These condition require one identifier pat-
tern to be set with TRIGger<m>:CAN:IMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:CAN:IMIN and TRIGger<m>:
CAN:IMAX on page 1496.
*RST: EQUal
TRIGger<m>:CAN:IMIN <IdPattern>
Specifies a message identifier pattern, or sets the the start value of an identifier range.
Parameters:
<IdPattern> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.