Remote control commands
R&S
®
RTE
1496User Manual 1326.1032.02 ─ 20
TRIGger<m>:CAN:IMAX <IdPatternTo>
Sets the the end value of an identifier range if TRIGger<m>:CAN:ICONdition is set
to INRange or OORange.
Parameters:
<IdPatternTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
TRIGger<m>:CAN:FDATa:FDF <FDFBit>
Specifies the CAN FD frame format. It corresponds to the EDL bit (extended data
length), which only exists in CAN FD format.
Parameters:
<FDFBit> ONE | ZERO | DC
ONE: CAN FD.
ZERO: CAN.
DC: don't care, the format is not relevant.
*RST: DC
TRIGger<m>:CAN:FDATa:BRS <BRSBit>
Sets the bit rate switch bit.
Parameters:
<BRSBit> ONE | ZERO | DC
ONE: the bit rate switches from the bit rate of the arbitration
phase to the faster data rate.
*RST: ONE
TRIGger<m>:CAN:FDATa:ESI <ESIBit>
Sets the error state indicator bit.
Parameters:
<ESIBit> ONE | ZERO | DC
DC: don't care, bit is nor relevant
*RST: DC
TRIGger<m>:CAN:DCONdition <DataOperator>
Sets the operator to set a specific data pattern or a data pattern range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
Protocols