Remote control commands
R&S
®
RTE
1515User Manual 1326.1032.02 ─ 20
Return values:
<NDBytes> Range: 1 to 64
Increment: 1
*RST: 1
Usage: Query only
SEARch:TRIGger:CAN:FDATa:DPOPerator <SearchName>,<DataPosOperator>
SEARch:TRIGger:CAN:FDATa:DPOPerator? <SearchName>
Sets the operator for the data position if DLC 9. You can define an exact position, or
a position range.
The setting is available in CAN FD option R&S RTE-K9.
Parameters:
<DataPosOperator> ANY | OFF | EQUal | GETHan | INRange | RANGe
ANY = OFF
The data position is not relevant for the search.
EQUal | GETHan
Equal, Greater or equal than. These conditions require one data
position to be set with SEARch:TRIGger:CAN:FDATa:
DPOSition.
INRange = RANGe
In range: Set the minimum and maximum value of the range with
SEARch:TRIGger:CAN:FDATa:DPOSition and SEARch:
TRIGger:CAN:FDATa:DPTO.
*RST:
ANY
Parameters for setting and query:
<SearchName> String with the search name
SEARch:TRIGger:CAN:FDATa:DPOSition <SearchName>,<DataPosition>
SEARch:TRIGger:CAN:FDATa:DPOSition? <SearchName>
Defines the first possible start position of the data pattern.
The setting is available in CAN FD option R&S RTE-K9.
Parameters:
<DataPosition> Range: 1 to 57
Increment: 1
*RST: 1
Parameters for setting and query:
<SearchName> String with the search name