Remote control commands
R&S
®
RTE
1516User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:CAN:FDATa:DPTO <SearchName>,<DataPositionTo>
SEARch:TRIGger:CAN:FDATa:DPTO? <SearchName>
Defines the last possible start position of the data pattern if the position operator
SEARch:TRIGger:CAN:FDATa:DPOPerator defines a range.
The setting is available in CAN FD option R&S RTE-K9.
Parameters:
<DataPositionTo> Range: 8 to 64
Increment: 1
*RST: 8
Parameters for setting and query:
<SearchName> String with the search name
SEARch:TRIGger:CAN:ACKerror
<SearchName>,<AckError>
SEARch:TRIGger:CAN:ACKerror? <SearchName>
Searches for acknowledgement errors. An acknowledgement error occurs when the
transmitter does not receive an acknowledgment - a dominant bit during the Ack Slot.
Parameters:
<AckError> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:BITSterror <SearchName>,<BitStuffError>
SEARch:TRIGger:CAN:BITSterror? <SearchName>
Searches for bit stuffing errors.
Parameters:
<BitStuffError> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:CAN:CRCerror <SearchName>,<ChecksumError>
SEARch:TRIGger:CAN:CRCerror? <SearchName>
Searches for errors in the Cyclic Redundancy Check.
Parameters:
<ChecksumError> ON | OFF
*RST: ON
Parameters for setting and query:
<SearchName>
Protocols