Remote control commands
R&S
®
RTE
1542User Manual 1326.1032.02 ─ 20
<n> *
Selects the frame.
Return values:
<IdentifierValue> To set the value format, use FORMat:BPATtern.
The values below range, increment and reset are decimal
values.
Range: 0 to 63
*RST: 0
Usage: Query only
BUS<m>:LIN:FRAMe<n>:IDPValue?
Returns the value of the identifier parity bits of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<IdtfParityValue> To set the value format, use FORMat:BPATtern.
The values below range, increment and reset are decimal
values.
Range: 0 to 3
*RST: 0
Usage: Query only
BUS<m>:LIN:FRAMe<n>:SYSTate?
Returns the state of the sync field for the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<SyncState> OK | STERror | SPERror | UVAL | NOEXists | INSufficient
STERror: start error
SPERror: stop error
UVAL: unexpected value
NOEXists: byte does not exist
INSufficient: the frame is not completely contained in the acqui-
sition. The decoded part of the frame is valid.
*RST: OK