Remote control commands
R&S
®
RTE
1595User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Coupling> ON | OFF
*RST: ON
BUS<m>:I2S:CLOCk:THReshold <SCLKThreshold>
Sets the threshold value for the clock line SCLK.
If BUS<m>:I2S:TCOupling is ON, the command sets the threshold for all lines.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SCLKThreshold> Range: -10 to 10
Increment: 1E-3
*RST: 1.6
Default unit: V
BUS<m>:I2S:WSELect:THReshold <WSThreshold>
Sets the threshold value for the word select and FSYNC lines.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<WSThreshold> Range: -10 to 10
Increment: 1E-3
*RST: 1.6
Default unit: V
BUS<m>:I2S:DATA:THReshold <SDATAThreshold>
Sets the threshold value for the data line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SDATAThreshold> Range: -10 to 10
Increment: 1E-3
*RST: 1.6
Default unit: V