Remote control commands
R&S
®
RTE
1613User Manual 1326.1032.02 ─ 20
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one pattern to
be set withTRIGger<m>:MILStd:CMD:SMIN.
INRange | OORange
In range/Out of range: set the minimum and maximum value of
the range with TRIGger<m>:MILStd:CMD:SMIN and
TRIGger<m>:MILStd:CMD:SMAX.
*RST: EQUal
TRIGger<m>:MILStd:CMD:SMAX
<SubaddrPattMax>
Sets the end value of the subaddress range if TRIGger<m>:MILStd:CMD:
SCONdition is set to INRange or OORange.
Parameters:
<SubaddrPattMax> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CMD:SMIN <SubaddrPattMin>
Specifies a subaddress or sets the the start value of a subaddress range.
Parameters:
<SubaddrPattMin> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:MILStd:CMD:TR <TRFlag>
Triggers on a transmission mode.
Parameters:
<TRFlag> ONE | ZERO | DC
ONE
Transmit direction.
ZERO
Receive direction.
DC
Either directions.
*RST: DC