Remote control commands
R&S
®
RTE
1744User Manual 1326.1032.02 ─ 20
BUS<m>:CMSB:FRAMe<n>:APPend
Creates an empty cell description and adds it to the end of the active frame descrip-
tion.
Suffix:
<m>
.
1..4
<n> *
Usage: Event
BUS<m>:CMSB:FRAMe<n>:CCOunt?
Returns the number of cells in the specified frame.
Suffix:
<m>
.
1..4
<n> *
Return values:
<Count>
Usage: Query only
BUS<m>:CMSB:FRAMe<n>:CELL<o>:NAME <CellName>
Enables the user to set a cell name within a frame. Names do not have to be unique,
they are just for user support.
Suffix:
<m>
.
1..4
<n> *
<o> *
Parameters:
<CellName>
BUS<m>:CMSB:FRAMe<n>:CELL<o>:BITCount <BitCount>
Sets the bit count of a cell, hence its length. Based upon the lengths of the previous
cells, this also defines the position of the cell start and end within a frame.
Suffix:
<m>
.
1..4
<n> *
<o> *