Remote control commands
R&S
®
RTE
1813User Manual 1326.1032.02 ─ 20
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater than,
Greater than or equal. These conditions require one address
value to be set using SEARch:TRIGger:USB:AMIN.
INRange | OORange
In range, out of range. These conditions require an address
range to be set using SEARch:TRIGger:USB:AMIN and
SEARch:TRIGger:USB:AMAX.
*RST: OFF
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:AMIN
<SearchName>,<Address>
SEARch:TRIGger:USB:AMIN? <SearchName>
Specifies an address, or sets the start value of an address range. The string parameter
does not accept the bit value X (don't care).
Parameters:
<Address>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:AMAX <SearchName>,<AddressTo>
SEARch:TRIGger:USB:AMAX? <SearchName>
Sets the the end value of an address range if TRIGger<m>:USB:ACONdition is set
to INRange or OORange. The string parameter does not accept the bit value X (don't
care).
Parameters:
<AddressTo>
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:USB:BITSterror <SearchName>,<BitstuffError>
SEARch:TRIGger:USB:BITSterror? <SearchName>
Defines, whether a search for any bitstuffing error shall be activated or not.
Parameters:
<BitstuffError> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName>
Protocols