Triggers
R&S
®
RTE
253User Manual 1326.1032.02 ─ 20
"Both"
Selects the rising as well as the falling edge. This option is not availa-
ble if the trigger source is the external trigger input.
Remote command:
TRIGger<m>:EDGE:SLOPe on page 1168
TRIGger<m>:ANEDge:SLOPe on page 1170
TRIGger<m>:SLEW:SLOPe on page 1181
Upper level
Sets the upper voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Remote command:
TRIGger<m>:LEVel<n>:SLEW:UPPer on page 1181
Lower level
Sets the lower voltage threshold. When the signal crosses this level, the slew rate
measurement starts or stops depending on the selected slope.
Remote command:
TRIGger<m>:LEVel<n>:SLEW:LOWer on page 1181
Range
Selects how the time limit for the slew rate is defined. The time measurement starts
when the signal crosses the first trigger level - the upper or lower level depending on
the selected slope. The measurement stops when the signal crosses the second level.
"Within"
Triggers on slew rates inside a given time range. The range is defined
by "Slew rate" and "±Delta".
"Outside"
Triggers on slew rates outside a given time range. The range defini-
tion is the same as for "Within" range.
"Shorter"
Triggers on slew rates shorter than the given "Slew rate" limit.
"Longer"
Triggers on slew rates longer than the given "Slew rate" limit.
Remote command:
TRIGger<m>:SLEW:RANGe on page 1182
Slew rate
For the ranges "Within" and "Outside", the slew rate defines the center of a range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", the slew rate defines the maximum and mini-
mum slew rate limits, respectively.
Remote command:
TRIGger<m>:SLEW:RATE on page 1182
±Delta
Defines a time range around the given slew rate.
Remote command:
TRIGger<m>:SLEW:DELTa on page 1183
Trigger types