Triggers
R&S
®
RTE
262User Manual 1326.1032.02 ─ 20
Remote command:
TRIGger<m>:LEVel<n>[:VALue] on page 1166
Signal polarity
Sets the polarity of the signal. Note that the sync pulse has the opposite polarity, for
example, a positive signal has a negative sync pulse.
Figure 6-2: Signal with positive polarity and tri-level sync pulse
Remote command:
TRIGger<m>:TV:POLarity on page 1194
Pulse type
Sets the type of the sync pulse, either bi-level sync pulse (used in SDTV signals), or tri-
level sync pulse (used in HDTV signals).
Figure 6-3: Bi-level (left) and tri-level (right) sync pulses
This setting is available for user-defined video signals if "Standard" is set to "Custom".
Remote command:
TRIGger<m>:TV:CUSTom:STYPe on page 1196
Line period
Sets the duration of a single video line, the time between two successive sync pulses.
Trigger types