Measurements
R&S
®
RTE
351User Manual 1326.1032.02 ─ 20
Meas. type Symbol Description/Result
Peak to peak X
PkPk
Peak-to-peak value of the waveform: the difference of maximum
and minimum values
X
Ampl
= X
Max
- X
Min
Mean X
Mean
Arithmetic average of the waveform voltage values
RMS X
RMS
RMS (root mean square, quadratic mean) of the waveform volt-
age values
Eval
N
i
Eval
RMS
ix
N
X
1
2
)(
1
σ
σ (S-dev)
σ
X
Standard deviation of the waveform samples
Eval
N
i
Mean
Eval
X
Xix
N
1
2
)(
1
1
Pos. overshoot R
Pos
Positive overshoot of a square wave, calculated from measure-
ment values High, Max, and Amplitude
%100
Amp
Ptop
V
VV
Ovr
Neg. overshoot R
Neg
Negative overshoot of a square wave, calculated from measure-
ment values Min, Low, and Amplitude
%100
Amp
Pbase
V
VV
Ovr
Cycle mean
The mean value of one cycle
Cycle RMS
The RMS (root mean square) value of one cycle
σ
Cycle σ (S-dev)
The standard deviation of one cycle
Trig. ProbeMeter