Measurements
R&S
®
RTE
367User Manual 1326.1032.02 ─ 20
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
The following figure shows a cursor measurement on a spectrum waveform:
If "Peak excursion" is 30 dB, the peaks 1 to 5 are found. If "Peak excursion" is 20 dB,
also the peaks 6 to 10 are found. The cursor position is on peak 6.
Remote command:
CURSor<m>:PEXCursion on page 1249
MEASurement<m>:SPECtrum:PEXCursion on page 1272
Max results
Sets the maximum number of measurement results that are listed in the result table.
Available for peak list and harmonic search measurements.
Remote command:
MEASurement<m>:SPECtrum:RESult<n>:COUNt on page 1273
Result mode
Selects the the way the measurement results are displayed. Available for peak list and
harmonic search measurements.