Loading...
Measurements
R&S
®
RTE
389User Manual 1326.1032.02 ─ 20
Enable (Track)
Enables the track of measurement results over time and displays the track waveform. It
is the graphical interpretation of all measurement values of a single acquisition.
The track is available for most amplitude/time measurements (except for High, Low,
Amplitude, Max, Min, Peak to peak, Mean, RMS, S-dev, Pos. and Neg. overshoot, and
Area), and for jitter measurements.
Enabling the track enables also the Continuous auto scale and Measure all events in
each acquisition.
Before you can enable the track, activate the appropriate measurement.
With option R&S RTE-K5 I²S audio signals, you can use the track for protocol mea-
surements on decoded audio buses, see Chapter 13.8.5, "Track", on page 651.
Remote command:
MEASurement<m>:TRACk[:STATe] on page 1301
Reset
Immediately resets the histogram, the long-term measurement and the statistics.
Stopping and starting the acquisition does not reset these analyses but only stops and
continues them.
To delete all results, waveforms and history, select "Display" menu > "Clear all".
Remote command:
MEASurement<m>:STATistics:RESet on page 1292
Analysis scale
The measurement scale of a long-term measurement diagram or measurement histo-
gram can be set automatically by the instrument, or manually.
Use automatic scaling if the measurement is running and you cannot see the expected
results.
"Continuous auto scale"
Performs an automatic scaling whenever the long-term waveform or
the histogram does not fit in the diagram during the measurement
period.
"Vertical"
Defines the vertical scaling per division for long-term measurement
period and the measurement histogram.
"Offset"
Defines an offset for the long-term measurement and the measure-
ment histogram.
Remote command:
MEASurement<m>:VERTical:CONT on page 1293
MEASurement<m>:VERTical:AUTO on page 1293
MEASurement<m>:VERTical:SCALe on page 1294
MEASurement<m>:VERTical:OFFSet on page 1293