452User Manual 1326.1032.02 ─ 20
Polarity, Range, Width, ±Delta
See trigger settings:
"Polarity" on page 246
While the width trigger can only analyze positive or negative polarity, searching for
a width is also possible for both polarities at the same time ("Either").
"Range" on page 246
"Width" on page 246
"±Delta" on page 247
Remote command:
SEARch:TRIGger:WIDTh:POLarity on page 1365
SEARch:TRIGger:WIDTh:RANGe on page 1365
SEARch:TRIGger:WIDTh:WIDTh on page 1366
SEARch:TRIGger:WIDTh:DELTa on page 1365
Trigger level
Sets the voltage level for trigger search. The value is used for all search event types
that require one trigger level. The search trigger level is search-specific, you can define
different levels in different searches for the same event.
Remote command:
SEARch:TRIGger:LEVel[:VALue] on page 1355
11.2.2.4 Runt
The runt search settings are the same as the runt trigger settings. To apply the trigger
settings to search, tap Copy 'A Trigger'.
The runt search is not available, if the search source is a digital channel.
Search setup