Protocol analysis
R&S
®
RTE
587User Manual 1326.1032.02 ─ 20
Standard
Selects the CAN standard: "CAN", "CAN FD", or "Any".
The setting is available in CAN FD option R&S RTE-K9.
Use "Any" to search for both CAN and CAN-FD frames. In this case, the search config-
uration provides all possible settings, for CAN as well as for CAN FD.
Remote command:
SEARch:TRIGger:CAN:FDATa:STANdard on page 1511
Frame type
Selects the frame type.
Remote frames are only available in the CAN protocol.
The frame types are the same as in the CAN trigger setup, see "Frame type"
on page 568.
Remote command:
SEARch:TRIGger:CAN:FTYPe on page 1511
ID type
Selects the length of the identifier.
"11 bit"
Identifier length of the CAN base frame format. The instrument trig-
gers on the sample point of the IDE bit.
"29 bit"
Identifier length of the CAN extended frame format. The instrument
triggers on the sample point of the RTR bit.
"Any"
The ID type is not relevant. If the trigger type is "Identifier + Data", set
the "ID type" to "Any" if you want to search only for data.
Remote command:
SEARch:TRIGger:CAN:ITYPe on page 1512
Identifier setup: Condition, Identifier min, Identifier max
The identifier setup consists mainly of the condition and one or two identifier patterns.
Additionally, ID type and frame type may qualify the identifier.
The identifier setup settings are the same as in the CAN trigger setup, see "Identifier
setup: Condition, Identifier min, Identifier max" on page 569.
CAN and CAN FD (option R&S