Protocol analysis
R&S
®
RTE
589User Manual 1326.1032.02 ─ 20
Figure 13-25: Identifier + Data search setup for CAN FD signals
Remote command:
SEARch:TRIGger:CAN:DCONdition on page 1513
SEARch:TRIGger:CAN:DMIN on page 1513
SEARch:TRIGger:CAN:DMAX on page 1514
SEARch:TRIGger:CAN:DLCCondition on page 1514
SEARch:TRIGger:CAN:DLC on page 1514
SEARch:RESult:CAN:FRAMe<m>:NDBYtes? on page 1514
Data position
The data position sets the location in the data field where the instrument looks for the
specified data pattern.
The setting is available in CAN FD option R&S RTE-K9.
The position can be defined if the data field of the frame is longer than 8 bytes - if DLC
≥9.
For details, see "Data position" on page 571.
Remote command:
SEARch:TRIGger:CAN:FDATa:DPOPerator on page 1515
SEARch:TRIGger:CAN:FDATa:DPOSition on page 1515
SEARch:TRIGger:CAN:FDATa:DPTO on page 1516
Error Condition
Selects the error type to be searched for. You can select one or more error types as
search condition.
The error types are the same as in the CAN trigger setup, see "Error conditions"
on page 572.
CAN and CAN FD (option R&S RTE-K3/K9)