Protocol analysis
R&S
®
RTE
606User Manual 1326.1032.02 ─ 20
Remote command:
TRIGger<m>:LIN:ICONdition on page 1533
TRIGger<m>:LIN:IMIN on page 1533
TRIGger<m>:LIN:IMAX on page 1533
Identifier OR setup: Monitor, Frame ID
Sets the trigger to a combination of up to four identifiers. Enter the patterns in the
"Frame ID" fields. In binary and hex format, characters 1, 0, and X (do not care) are
allowed. For each identifier pattern to be triggered on, enable "Monitor".
Remote command:
TRIGger<m>:LIN:IDOR<n>:ENABle on page 1535
TRIGger<m>:LIN:IDOR<n>[:VALue] on page 1536
Data setup: Data length, Transfer, Condition, Data min, Data max
The data setup consists of the transfer direction, the number of bytes, the condition,
and one or two data patterns.
"Transfer"
Sets the byte order (endianness) of the data transfer. With "Big
endian", the data is analyzed and evaluated in the order of reception.
With "Little endian", the instrument reads the complete data, reverses
the byte order of the data, and compares it with the specified data
word.
According to the standard, LIN data is transmitted in little endian
transfer order. The "Little endian" setting allows you to enter the
required data word directly into "Data min", and the instrument trig-
gers correctly.
LIN (option R&S RTE-K3)