Protocol analysis
R&S
®
RTE
664User Manual 1326.1032.02 ─ 20
Trigger Type
Selects the trigger type for MIL-STD-1553 analysis.
"Sync"
Triggers on a sync impulse.
"Word"
Triggers on the selected word type.
"Data Word"
Triggers on a specified data word or data word range.
"Command/ Status Word"
Triggers on a specified command word or on a status word.
"Command Word"
Triggers on a specified command word.
"Status Word"
Triggers on a specified status word.
"Error Condi-
tion"
Triggers on any combination of protocol errors.
Remote command:
TRIGger<m>:MILStd:TYPE on page 1610
Sync Type / Word Type
Triggers on a sync impulse or word type. You can select to trigger on "Command/
Status", on "All" or on "Data" sync pulses / word types.
Remote command:
TRIGger<m>:MILStd:TPSPecifier on page 1619
Remote Terminal Address
The RTA setup consists of the condition and one or two RTA patterns.