Protocol analysis
R&S
®
RTE
683User Manual 1326.1032.02 ─ 20
Label + Data
Sets the trigger on a defined word format. You can define the label, the data and the
SDI / SSM bits separately.
Label setup: Condition, Label Min, Label Max ← Label + Data
The label setup consists of the condition and one or two label patterns.
"Condition"
Defines the operator to set a specific label ("Equal" or "Not equal") or
a label range.
"Label Min"
Defines the bit pattern of the label.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.5, "Bit pattern editor", on page 521.
"Label Max"
The second label pattern is required to specify a range with condi-
tions "In range" and "Out of range".
Remote command:
TRIGger<m>:ARINc:LABel:CONDition on page 1637
TRIGger<m>:ARINc:LABel:MIN on page 1637
TRIGger<m>:ARINc:LABel:MAX on page 1638
Data setup: Condition, Data Min, Data Max ← Label + Data
The data setup consists of the condition and one or two data patterns.
"Condition"
Defines the operator to set a specific data ("Equal" or "Not equal") or
a data range.
"Data Min"
Defines the bit pattern of the data.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.5, "Bit pattern editor", on page 521.
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
ARINC 429 (option R&S RTE-K7)