10 waveform acquisitions are performed. The most recent acquisition is displayed
in the diagram.
5. In the [ANALYSIS] area, press the [HISTORY] key.
The quick access "History" dialog box appears, and the history mode is on.
6. Tap "Play".
The 10 stored waveforms are displayed one after the other, but very fast.
7. In the "Current acq." field, enter -4 to display the sixth waveform, counted from
acquisition start. The latest acquisition has the number 0, the oldest has -9.
8.
Tap
to open the "History" setup dialog box.
9. Enter -2 in the "Start acq" field. Tap "Newest" to enter 0 in the "Stop acq" field.
Thus the three latest acquisitions are displayed.
10. In the "Replay time per acq." field, enter 1 s to display each waveform for one sec-
ond.
11. Enable the "Auto repeat" option to see the three waveforms repeatedly.
12. Tap "Play".
The currently displayed waveform is indicated in the "Current acq." field.
13. Close the "History" dialog box so you can see the waveform better.
14. Tap "Running" to stop the display.
During running display, the "Play" is labeled "Running".
15. Close the quick access "History" dialog box.
The history mode is disabled. The [HISTORY] key is no longer illuminated.