Protocol analysis
R&S
®
RTE
869User Manual 1326.1032.02 ─ 20
"Data pattern"
Sets the trigger to a defined data pattern or pattern range.
"NULL frame"
Sets the trigger to a null frame, a frame without usable data.
"Time Code"
Sets the trigger to a time-code control code. You can define the data
pattern of the time code to be triggered on.
"Errors"
Triggers on an enabled error type.
Remote command:
TRIGger<m>:SWIRe:TYPE on page 1861
Trigger Control Type
Triggers on a specific control type character.
"ANY"
Any control type character
"FCT"
Flow control token character
"EOP"
Normal end of packet character
"EEP"
Error end of packet character
Remote command:
TRIGger<m>:SWIRe:CTYPe on page 1859
Data (Time Code)
Sets the specified data type for the time code to be triggered on. The trigger data setup
consists of a comparison condition and one or two data patterns.
In binary format, use the following characters: 1; 0; or X (do not care). To enter the pat-
tern in other formats use the bit pattern editor, see Chapter 13.1.5, "Bit pattern editor",
on page 521.
"Condition"
Sets the comparison condition to a specific value or a range.
"Data Min/
Data"
Specifies the value or sets the start value of a range.
SpaceWire (option R&S RTE-K65)