Power analysis (option R&S RTE-K31)
R&S
®
RTE
1001User Manual 1326.1032.02 ─ 20
●
The result box displays the numeric measurement results for the enabled measure-
ment parameters in dependence of the energy and the power. To switch the dis-
play, tap "Energy" or "Power" accordingly.
For a detailed description, see Table 16-2.
To measure and display the switching loss, the instrument uses the following measure-
ments and cursors:
●
"P3"... "P8": "Meas 3" ... "Meas 8" to measure the voltage
●
"M4" Math 4 to calculate the power
●
"C3" Cursor 3 to define time points "t
0
" and "t
1
"
●
"C4" Cursor 4 to define time points "t
2
" and "t
3
"
The used resources are listed in the "Details" tab. See also: Chapter 16.2.3, "Details
tab", on page 966.
"Switching Loss" is a statistical evaluation that is reset only if the measurement setup
is changed or you reset the statistics.
The switching loss phases that can be defined during the measurement are shown in
Figure 16-4 and described in Table 16-4.
Figure 16-4: Switching loss phases
Power path