Remote control commands
R&S
®
RTE
1150User Manual 1326.1032.02 ─ 20
Parameters:
<ExtShuntMeasRg> V450m | V045m
V450m: 450 mV
V045m: 45 mV
*RST: V450m
Usage: Asynchronous command
ZVC:Z<m>:I<n>:SKEW <DeskewOffset>
Sets the skew offset value for the current probe channel. This is a delay value, that is
known from the circuit specifics but cannot be compensated by the instrument auto-
matically.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<DeskewOffset> Range: -6E-6 to 6E-6
Increment: 200E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
ZVC:Z<m>:I<n>[:STATe] <State>
Enables the corresponding current channel of the probe.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
ZVC:Z<m>:V<n>:BANDwidth?
Queries the bandwidth of the voltage channel. You can set the probe bandwidth with
ZVC:BANDwidth.