Protocol analysis
R&S
®
RTE
575User Manual 1326.1032.02 ─ 20
Example: Triggering on the seventh or later data byte
The CAN FD frame has 12 or more data bytes, where the value of the seventh or later
data byte should be 17.
●
Set "DLC ≥ 9".
●
Set the data pattern: "= XX XX XX XX XX XX 17 XX".
Figure 13-17: Trigger setup to trigger on the 7th or later data byte with value = 17
CAN and CAN FD (option R&S RTE-K3/K9)