Protocol analysis
R&S
®
RTE
576User Manual 1326.1032.02 ─ 20
Figure 13-18: Data type trigger on data byte with the 7th or later byte value = 17
Example: Triggering on a data byte at a given position
The CAN FD frame has 12 or more data bytes. the trigger is set at the 8th data byte
starting at the 4th data byte or later, with data byte value = 17.
●
Set "DLC ≥ 9".
●
Set the data pattern: "= XX XX XX XX XX XX XX 17 ".
●
Set the position of the data pattern: "In range", "4" to "12".
CAN and CAN FD (option R&S RTE-K3/K9)