19.4.1.2 Normalization
Open [App Cockpit]> "Analysis" tab > "Proven Probe / Cable" > "Normalize setup"
1. Connect the cables to the pulse source R&S RTP-B7 and to the input channels.
The probe or cable to be measured is not connected.
Use only high-quality double-shielded RF cables. Two suitable cables are delivered
with the pulse source option R&S RTP-B7.
Figure 19-1: Normalization settings
2. Start measurements with "Normalize".
The basic setup is characterized.
Device
Selects the input channel that is connected to the connector of the pulse source
R&S RTP-B7 during normalization, and to the measured probe or cable during mea-
surements.
Reference
Selects the input channel that is connected to Out connector of the pulse source.
Proven probe, proven cable