Network operation and remote control
R&S
®
RTP
1110User Manual 1337.9952.02 ─ 12
Figure 21-1: Overview of the status registers hierarchy
All status registers have the same internal structure.
21.5.4.2 Contents of the status registers
STATus:OPERation register
In the CONDition part, this register contains information on which actions the instru-
ment is executing. In the EVENt part, it contains information on which actions the
instrument has executed since the last reading.
It can be read using the commands STATus:OPERation:CONDition? or
STATus:OPERation[:EVENt]?, see also Chapter 22.25.1, "STATus:OPERation reg-
ister", on page 2306.