Remote control commands
R&S
®
RTP
1152User Manual 1337.9952.02 ─ 12
Determines whether the contents of the ENABle registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
<Action> 0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL <Number>
Recall
Loads the instrument settings from an intermediate memory identified by the specified
number. The instrument settings can be stored to this memory using the command
*SAV with the associated number.
*RST
Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
Usage:
Setting only
*SAV <Number>
Save
Stores the current instrument settings under the specified number in an intermediate
memory. The settings can be recalled using the command *RCL with the associated
number.
*SRE <Contents>
Service request enable
Sets the service request enable register to the indicated value. This command deter-
mines under which conditions a service request is triggered.
Parameters:
<Contents> Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range: 0 to 255
Common commands