Remote control commands
R&S
®
RTP
1156User Manual 1337.9952.02 ─ 12
SYSTem:DISPlay:UPDate <Enable>
Defines whether the display is updated while the instrument is in the remote state. If
the display is switched off, the normal GUI is replaced by a static image while the
instrument is in the remote state. Switching off the display can speed up the measure-
ment. This is the recommended state.
See also: Chapter 21.5.3.2, "Using the display during remote control", on page 1109
Parameters:
<Enable> ON | 1: Display is shown and updated during remote control
OFF | 0: Display shows static image during remote control
Example:
SYSTem:DISPlay:UPDate 1
Switch on the display update.
SYSTem:KLOCk
<Enable>
Locks or unlocks the local controls of the instrument. This includes the front panel
keys, the keyboard, or other local interfaces. except for the "View" button on the dis-
play.
Parameters:
<Enable> ON | 1: Locks the local keys
OFF | 0: Keys are unlocked
Usage: SCPI confirmed
GPIB:ADDRess <Address>
Sets the GPIB address of the instrument. Changing the address has major effects on
the communication to the remote computer.
Parameters:
<Address> Range: 0 to 30
Increment: 1
*RST: 20
GPIB:TERMinator <Terminator>
Specifies the symbol that is used as a terminator in GPIB communication.
Parameters:
<Terminator> LFEoi | EOI
*RST: EOI
SYSTem:DISPlay:MESSage:STATe <DispMessSt>
Enables and disables the display of an additonal text in remote control.
To define the text, use SYSTem:DISPlay:MESSage[:TEXT].
General remote settings