Remote control commands
R&S
®
RTP
1203User Manual 1337.9952.02 ─ 12
CHANnel<m>:CPLing <Channel>, <State>
CHANnel<m>:CPLing? <Channel>
Sets the vertical settings of the coupled channel to the values of the active channel.
Suffix:
<m>
.
1..4
Active channel
Parameters:
<State> OFF | ON
Parameters for setting and query:
<Channel> CHAN1 | C1 | CHAN2 | C2 | CHAN3 | C2 | CHAN4 | C4
CHAN1 = C1, CHAN2 = C2, CHAN3 = C2, CHAN4 = C4
Input channel that takes the vertical settings of CHANnel<m>.
Usage: Asynchronous command
CHANnel<m>:IMPedance <Impedance>
Sets the impedance of the channel for power calculations and measurements.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Impedance> Range: 0.1 to 100E+3
Increment: 1
*RST: 50
Default unit: Ohm
Usage: Asynchronous command
CHANnel<m>:OVERload <Overload>
Retrieves the overload status of the specified channel from the status bit. When the
overload problem is solved, the command resets the status bit.
The overload status is returned asynchronously. Therefore, it is not possible to assign
an overload to a specific acquisition during continuous acquisition. This assignment is
only possible for a single acquisition.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Overload> ON | OFF
Use OFF to reset the overload status bit.
*RST: OFF
Acquisition and setup