Remote control commands
R&S
®
RTP
1216User Manual 1337.9952.02 ─ 12
TRPRobe:SETup:ZAXV <ExtAttRTZA15>
PROBe<m>:SETup:ZAXV <ExtAttRTZA15>
If you use the external attenuator R&S RT-ZA15 together with one of the differential
active probes R&S RT-ZD10/20/30, enable it to include the external attenuation in the
measurements.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ExtAttRTZA15> ON | OFF
*RST: OFF
Usage: Asynchronous command
TRPRobe:SETup:ACCoupling
<ProbeCouplingAC>
PROBe<m>:SETup:ACCoupling <ProbeCouplingAC>
Enables AC coupling in the R&S RT-ZPR20 probe, which removes DC and very low-
frequency components. The R&S RT-ZPR20 probe requires 50 input termination, for
which the channel AC coupling is not available. The probe setting allows AC coupling
also at 50 inputs.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ProbeCouplingAC> ON | OFF
*RST: OFF
Usage: Asynchronous command
PROBe<m>:SETup:ADVanced:PMToffset
Sets the measured ProbeMeter value as offset. Make sure that the ProbeMeter is
active before you use this command.
Only available for power rail probes R&S RT-ZPR.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
PROBe<m>:SETup:ADVanced:RANGe <ProbeRange>
Sets the voltage range of a R&S RT-ZHD probe.