Remote control commands
R&S
®
RTP
1217User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ProbeRange> AUTO | MHIGh | MLOW
AUTO
The voltage range is set with CHANnel<m>:SCALe.
MHIGh
Sets the higher voltage range of the connected probe. To query
the value, use PROBe<m>:SETup:ATTenuation[:AUTO]?.
MLOW
Sets the lower voltage range of the connected probe. To query
the value, use PROBe<m>:SETup:ATTenuation[:AUTO]?.
*RST: AUTO
Usage: Asynchronous command
TRPRobe:SETup:ADVanced:FILTer <BwFilterSt>
PROBe<m>:SETup:ADVanced:FILTer <BwFilterSt>
Enables the lowpass filter in the probe control box.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<BwFilterSt> ON | OFF
*RST: OFF
Usage: Asynchronous command
TRPRobe:SETup:ADVanced:AUDioverload <AudibOvrrg>
PROBe<m>:SETup:ADVanced:AUDioverload <AudibOvrrg>
Activates the acustic overrange warning in the probe control box.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Sound> ON | OFF
*RST: OFF
Usage: Asynchronous command
Acquisition and setup