Remote control commands
R&S
®
RTP
1225User Manual 1337.9952.02 ─ 12
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<PredefinedProbe> ZC10 | ZC20 | ZC30 | ZD01A100 | ZD01A1000 | ZZ80 | FREE |
ZS10L | ZD02 | ZD08 | ZC02100 | ZC021000 | ZC03 |
ZD002A10 | ZD002A100 | ZD003A20 | ZD003A200 | ZC3110 |
ZC311 | ZC3101
ZC10 | ZC20 | ZC30 | ZZ80 | ZS10L | ZD02 | ZD08 | ZC03
Type of the probe
ZD01A100 | ZD01A1000 | ZD002A10 | ZD002A100 |
ZD003A20 | ZD003A200
High voltage differential probes R&S RT-ZD0xx, attenuation ratio
according to the setting on the probe.
A10 = 10:1
A20 = 20:1
A100 = 100:1
A200 = 200:1
A1000 = 1000:1
ZC02100 | ZC021000
Current probes 100 A or 1000 A according to the setting on the
probe.
FREE
Any other probe that is not recognized by the instrument.
*RST:
FREE
Usage: Asynchronous command
PROBe<m>:SETup:OFFSet:TOMean
Performs an automatic compensation for a DC component of the specified input signal
using the result of a background mean measurement.
Suffix:
<m>
.
1..4
Selects the input channel.
Usage: Event
Asynchronous command
22.8.7.8 Current probes
To set up R&S RT-ZC10 and R&S RT-ZC20, use PROBe<m>:SETup:ATTenuation:
DEFProbe.
TRPRobe:SETup:DEGauss...........................................................................................1226
PROBe<m>:SETup:DEGauss....................................................................................... 1226
PROBe<m>:SETup:OFFSet:STPRobe...........................................................................1226
PROBe<m>:SETup:OFFSet:ZADJust.............................................................................1226
Acquisition and setup