EasyManua.ls Logo

Rohde & Schwarz RTP Series - Page 1236

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1236User Manual 1337.9952.02 ─ 12
ZVC:Z<m>:I<n>:SHUNt:MAXVoltage <ExtShuntMeasRg>
Sets the maximum voltage for the external shunt.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<ExtShuntMeasRg> V450m | V045m
V450m: 450 mV
V045m: 45 mV
*RST: V450m
Usage: Asynchronous command
ZVC:Z<m>:I<n>:SKEW <DeskewOffset>
Sets the skew offset value for the current probe channel. This is a delay value, that is
known from the circuit specifics but cannot be compensated by the instrument auto-
matically.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<DeskewOffset> Range: -6E-6 to 6E-6
Increment: 200E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
ZVC:Z<m>:I<n>[:STATe] <State>
Enables the corresponding current channel of the probe.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command