Remote control commands
R&S
®
RTP
1242User Manual 1337.9952.02 ─ 12
Usage: Query only
Asynchronous command
ACQuire:ZRESolution?
Returns the current resolution of the R&S RT-ZVCmulti-channel power probe channel.
Return values:
<ZUIResolution> Range: 1E-15 to 0.5
Increment: 10E-12
*RST: 500E-6
Default unit: s
Usage: Query only
Asynchronous command
ZVC:Z<m>:I<n>:DATA:HEADer?
ZVC:Z<m>:V<n>:DATA:HEADer?
Return the header of amperemeter and voltmeter waveform data.
Table 22-2: Header data
Position Meaning Example
1 Start time in s -0.0001
2 Stop time in s 0.0001
3 Number of values in one waveform (record length) 1000
4 Number of values per sample interval. For most
waveforms the result is 1, for peak detect and enve-
lope waveforms it is 2. If the number is 2, the num-
ber of returned values is twice the number of values
(record length)..
1
Suffix:
<m>
.
1..2
<n> 1..4
Example:
ZVC:Z2:V2:DATA:HEADer?
<-- -0.0001,0.0001,1000,1
Usage: Query only
ZVC:Z<m>:I<n>:DATA[:VALues]?
ZVC:Z<m>:V<n>:DATA[:VALues]?
Return the waveform data of the amperemeter and voltmeter for transmission from the
instrument to the controlling computer. The data can be used in MATLAB, for example.
Set the export format to ASCii or REAL,32 using FORMat[:DATA]. Export of raw data
is not supported.