Remote control commands
R&S
®
RTP
1246User Manual 1337.9952.02 ─ 12
VertScale
P
= VertScale
N
= VertScale
CM
= VertScale
Diff
/2
You can disable the scale coupling and set an individual scale for each waveform.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<ScaleCoupling> ON | OFF
*RST: ON
Usage: Asynchronous command
DIFFerential<m>:COMMon:OFFSet <Offset>
DIFFerential<m>:DIFFerential:OFFSet <Offset>
Set the offset of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Offset> Offset value
Range: -1 to 1
Increment: 0.01
*RST: 0
Default unit: V
Usage: Asynchronous command
DIFFerential<m>:COMMon:POSition <Position>
DIFFerential<m>:DIFFerential:POSition <Position>
Set the vertical position of differential and common mode waveforms.
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Position> Position value
Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command
Acquisition and setup