Remote control commands
R&S
®
RTP
1248User Manual 1337.9952.02 ─ 12
CHANnel<m>:SKEW:MANual <ManualCompens>
If enabled, the skew offset value (CHANnel<m>:SKEW:TIME) is used for compensa-
tion. This improves horizontal and trigger accuracy.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualCompens> ON | OFF
*RST: ON
Usage: Asynchronous command
CHANnel<m>:SKEW:TIME
<Offset>
Sets an delay value, that is known from the circuit specifics but cannot be compensa-
ted by the instrument automatically. It affects only the selected input channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Offset> Range: -100E-9 to 100E-9
Increment: 10E-15
*RST: 0
Default unit: s
Usage: Asynchronous command
PROBe<m>:SKEState <ProbeSkew>
If enabled, the skew of all connected active probes is measured, displayed, and used
for deskewing.
Suffix:
<m>
.
1..4
The suffix is irrelevant. The setting affects all active channels.
Parameters:
<ProbeSkew> ON | OFF
*RST: OFF
Usage: Asynchronous command
CHANnel<m>:SKEW:EXTended <ExtendSkew>
Enables a higher skew value range that can be set with CHANnel<m>:SKEW:TIME
on page 1248. The skew extension reduces the maximum number of acquisitions in
the memory.