Remote control commands
R&S
®
RTP
1299User Manual 1337.9952.02 ─ 12
TRIGger<m>:TV:CUSTom:LDURation <LinePeriod>
Sets the duration of a line, the time between two successive sync pulses. Only relevant
if TRIGger<m>:TV:STANdard is set to CUSTom.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<LinePeriod> Range: 1E-6 to 500E-6
Increment: 100E-9
*RST: 64E-6
Default unit: s
Usage: Asynchronous command
TRIGger<m>:TV:CUSTom:STYPe <SyncPulseType>
Sets the type of the sync pulse. Only relevant if TRIGger<m>:TV:STANdard is set to
CUSTom.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<SyncPulseType> BILevel | TRILevel
BILevel
Bi-level sync pulse, usually used in SDTV signals
TRILevel
Tri-level sync pulse, used in HDTV signals
*RST: BILevel
Usage: Asynchronous command
TRIGger<m>:TV:CUSTom:SDURation <SyncPulseDurat>
Sets the width of the sync pulse. Only relevant if TRIGger<m>:TV:STANdard is set to
CUSTom.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<SyncPulseDurat> Range: 100E-9 to 100E-6
Increment: 100E-9
*RST: 4.7E-6
Default unit: s
Usage: Asynchronous command