Remote control commands
R&S
®
RTP
1354User Manual 1337.9952.02 ─ 12
CURSor<m>:XCOupling <Coupling>
Defines the positioning mode of the vertical cursor.
Suffix:
<m>
.
1..4
Selects the cursor set.
Parameters:
<Coupling> ON | OFF
ON
Moving one cursor line moves the other cursor line too. The cur-
sor lines always remain a fixed distance.
OFF
Each cursor line is positioned independently.
*RST: OFF
Usage: Asynchronous command
CURSor<m>:Y1Position <YPosition1>
Defines the position of the lower horizontal cursor line.
If CURSor<m>:TRACking[:STATe] is enabled, the query returns the measurement
result - the lower vertical value of the waveform.
Suffix:
<m>
.
1..4
Selects the cursor set.
Parameters:
<YPosition1> Range: -50 to 50
Increment: 0.01
*RST: 0
Default unit: The unit depends on the type of the waveform.
Usage: Asynchronous command
CURSor<m>:Y2Position <YPosition2>
Defines the position of the upper horizontal cursor line.
If CURSor<m>:TRACking[:STATe] is enabled, the query returns the measurement
result - the upper vertical value of the waveform.
Suffix:
<m>
.
1..4
Selects the cursor set.
Cursor measurements