MEASurement<m>:AMPTime:LCHeck<n>:LOWer:MARGin............................................. 1380
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:MARGin.............................................. 1380
MEASurement<m>:ENVSelect <EnvelopeCurve>
The command is only relevant for measurements on envelope waveforms. It selects
the envelope to be used for measurement.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<EnvelopeCurve> MIN | MAX | BOTH
MIN: measures on the lower envelope
MAX: measures on the upper envelope
BOTH: the envelope is ignored and the waveform measured as
usual
*RST: BOTH
Usage: Asynchronous command
MEASurement<m>:DETThreshold <SignDetectThres>
Defines the value above which measurement results are displayed. Values beneath the
threshold are considered to be noise and they are ignored.
Suffix:
<m>
.
1..10
irrelevant
Parameters:
<SignDetectThres> Range: 0 to 50
Increment: 1
*RST: 5
Default unit: %
Usage: Asynchronous command
MEASurement<m>:AMPTime:ALEVel <AreaLevel>
Defines the reference level used to integrate the waveform.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Automatic measurements