Remote control commands
R&S
®
RTP
1397User Manual 1337.9952.02 ─ 12
Parameters:
<State> ON | OFF
Parameters for setting and query:
<MeasType> PLISt
Only for peaklist measurements. For other measurements, this
command returns an error.
Example: Display labels.
MEAS:RES:SHL PLIS,ON
MEASurement<m>:SPECtrum:LCHeck<n>:VALid
<ValidRange>
Enables or disables limit checking for spectrum measurements in the specified mea-
surement channel.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
<n> 1..15
Number of the spectrum measurement, see Table 22-8.
Parameters:
<ValidRange> ILIMit | ULIMit | LLIMit | OLIMit
ILIMit
Inside (within) limit; between the upper and lower limit values
ULIMit
Upper limit; above the upper limit value
LLIMit
Lower limit; below the lower limit value
OLIMit
Outside limit; above the upper limit or below the lower limit val-
ues
*RST: ILIMit
Usage: Asynchronous command
MEASurement<m>:SPECtrum:LCHeck<n>:LOWer:LIMit <LowerLimit>
MEASurement<m>:SPECtrum:LCHeck<n>:UPPer:LIMit <UpperLimit>
Define the lower and upper limits for the limit check, respectively. The valid range is
defined using the MEASurement<m>:SPECtrum:LCHeck<n>:VALid command.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
<n> 1..15
Number of the spectrum measurement, see Table 22-8.
Automatic measurements