EasyManua.ls Logo

Rohde & Schwarz RTP Series - Page 1405

Rohde & Schwarz RTP Series
2403 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTP
1405User Manual 1337.9952.02 ─ 12
<MeasType>
parameter value
(remote control)
Measurement
(manual opera-
tion)
Description/Result Suffix
<n>
UPEakvalue Upper peak value Signal value at the maximum sample value in the
upper half of the histogram
6
LPEakvalue Lower peak value Signal value at the maximum sample value in the
lower half of the histogram
7
HMAXimum Maximum Highest signal value with a probability > 0 8
HMINimum Minimum Lowest signal value with a probability > 0 9
MEDian Median Signal value for which half the samples lie above,
the other half below in the histogram
10
MAXMin Max - Min Range of signal values with a probability > 0 11
HMEan Mean Weighted arithmetic average of the histogram 12
HSTDdev σ (S-dev) Standard deviation of the sample numbers 13
M1STddev Mean ±σ Range between (mean value + standard deviation)
and (mean value - standard deviation)
14
M2STddev Mean ±2*σ Range between (mean value + 3 * standard devia-
tion) and (mean value - 2 * standard deviation)
15
M3STddev Mean ±3*σ Range between (mean value + 3 * standard devia-
tion) and (mean value - 2 * standard deviation)
16
MKPositive Marker + Probabil-
ity %
Marker value (according to the selected probability
domain marker type) plus the defined limit.
Note that the value is restricted to the histogram
range.
17
MKNegative Marker - Probability
%
Marker value (according to the selected probability
domain marker type) minus the defined limit.
Note that the value is restricted to the histogram
range.
18
MEASurement<m>:HISTogram:SELect.......................................................................... 1405
MEASurement<m>:HISTogram:PROBability:TYPE..........................................................1406
MEASurement<m>:HISTogram:PROBability:LIMit........................................................... 1406
MEASurement<m>:HISTogram:LCHeck<n>:VALid.......................................................... 1407
MEASurement<m>:HISTogram:LCHeck<n>:LOWer:LIMit.................................................1407
MEASurement<m>:HISTogram:LCHeck<n>:UPPer:LIMit................................................. 1407
MEASurement<m>:HISTogram:LCHeck<n>:LOWer:MARGin........................................... 1408
MEASurement<m>:HISTogram:LCHeck<n>:UPPer:MARGin............................................ 1408
MEASurement<m>:HISTogram:SELect <HistgRef>
Selects the histogram on which the measurement is based.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.