Remote control commands
R&S
®
RTP
1425User Manual 1337.9952.02 ─ 12
MEASurement<m>:GATE:CURSor <Cursorset>
Selects the cursor set to be used for measurement gating. The gate area is defined by
the cursor lines.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<Cursorset> CURSOR1 | CURSor1 | CURSOR2 | CURSor2 | CURSOR3 |
CURSor3 | CURSOR4 | CURSor4
CURSOR1 = CURSor1, CURSOR2 = CURSor2, CURSOR3 =
CURSor3, CURSOR4 = CURSor4
*RST: CURSOR1
Usage: Asynchronous command
MEASurement<m>:GATE:CCOupling <CursorCoupling>
Enables the cursor coupling for automatic measurements.
Select the cursor set to be used with MEASurement<m>:GATE:CURSor.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<CursorCoupling> ON | OFF
*RST: OFF
Usage: Asynchronous command
MEASurement<m>:GATE:ZCOupling <ZoomCoupling>
If enabled, the gate area is defined identically to the zoom area for the zoom diagram.
If enabled, define the zoom area to be used as gate with MEASurement<m>:GATE:
ZDIagram.
Suffix:
<m>
.
1..10
See "Selection of the measurement group: MEASurement<m>"
on page 1359.
Parameters:
<ZoomCoupling> ON | OFF
*RST: OFF
Usage: Asynchronous command