Remote control commands
R&S
®
RTP
1478User Manual 1337.9952.02 ─ 12
MTESt:EYEMask:HPERiod <MaskTestName>,<InterpattLength>
MTESt:EYEMask:HPERiod? <MaskTestName>
Defines the time distance between the shape centers if MTESt:EYEMask:MSKLeft
and/or MTESt:EYEMask:MSKRight are ON.
Parameters:
<InterpattLength> Range: 0 to 100
Increment: 0.01
*RST: 0.5
Default unit: s
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:EYEMask:MSKTop <MaskTestName>,<Add>
MTESt:EYEMask:MSKTop? <MaskTestName>
MTESt:EYEMask:MSKBottom <MaskTestName>,<Add>
MTESt:EYEMask:MSKBottom? <MaskTestName>
Enable the upper (top) and lower (bottom) mask region, respectively.
Parameters:
<Add> ON | OFF
*RST: OFF
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:EYEMask:TOFFset <MaskTestName>,<RectDistTop>
MTESt:EYEMask:TOFFset? <MaskTestName>
MTESt:EYEMask:BOFFset <MaskTestName>,<RectDistBtom>
MTESt:EYEMask:BOFFset? <MaskTestName>
Voltage distance from the eye shape center that limit the upper (TOFFset) and lower
(BOFFset) regions.
Parameters:
<RectDistBtom> Range: 0 to 100
Increment: 0.01
*RST: 0.5
Default unit: V
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
Mask testing