Remote control commands
R&S
®
RTP
1515User Manual 1337.9952.02 ─ 12
Parameters:
<ResultListLimit> Range: 1 to 1000
Increment: 1
*RST: 100
Parameters for setting and query:
<SearchName> Search definition
Example: See Chapter 22.3.4.1, "Searching for a pulse of specified width",
on page 1132
Usage: Asynchronous command
SEARch:RESult:SHOW
<SearchName>,<ShowResultTable>
SEARch:RESult:SHOW? <SearchName>
Displays or hides the search result table.
Parameters:
<ShowResultTable> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
SEARch:RESult:SORT:ASCending <SearchName>,<SortAscending>
SEARch:RESult:SORT:ASCending? <SearchName>
If enabled, the results are listed in asscending order, i.e. the smallest value at the top.
Parameters:
<SortAscending> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
SEARch:RESult:SORT[:MODE] <SearchName>,<SortMode>
SEARch:RESult:SORT[:MODE]? <SearchName>
Sorts the search result table by x-value position or value of the result.
Parameters:
<SortMode> POSition | VALue
POSition
Sorts the search result table by the x-value position.
VALue
Sorts the search result table by the value of the result.