Remote control commands
R&S
®
RTP
1534User Manual 1337.9952.02 ─ 12
Parameters:
<IncHorValues> ON | OFF
*RST: OFF
Example: See:
"Exporting interleaved x/y data of a single waveform to CSV file"
on page 1135
"Exporting interleaved x/y data of a zoom to CSV file"
on page 1136
Usage: Asynchronous command
EXPort:WAVeform:RAW
<RawValues>
Enables the export of raw sample data, and sets the data format to integer 8 bit. In
high definition acquisition mode, the data format is integer 16 bit. Fot INT16, you can
set the byte order using the FORMat:BORDer command.
The raw format reduces the file size but changes also the precision of the values.
Currently, the setting is not available for the export of digital channel data and data of
R&S RT-ZVC channels.
The raw export of interleaved X/Y values is not supported.
Parameters:
<RawValues> ON | OFF
*RST: OFF
Example: See:
"Exporting raw data of a single waveform to BIN file"
on page 1134
"Exporting raw data of a measurement gate to BIN file"
on page 1134
Usage: Asynchronous command
EXPort:WAVeform:DISPlayoff <FastExport>
Enables or disables the display update during an Nx Single acquisition.
Parameters:
<FastExport> ON | OFF
ON: Disables the display update for maximum export speed.
OFF: Enables the display update. The export is slower.
*RST: OFF
Usage: Asynchronous command