Acquisition and waveform setup
R&S
®
RTP
157User Manual 1337.9952.02 ─ 12
The waveform values are recorded with 16-bit word length, except for peak detect dec-
imation (2 values with 8 bit).
Vertical system
The current bandwidth is shown in the channel label.
In the "Vertical" > "Bandwidth" dialog (C<x>), the "Bandwidth" setting is not available
because the bandwidth is set by the high definition filter.
The minimum vertical scale is 1 mV/div instead of 2 mV/div in normal mode.
Digital filter
The digital filter settings are set automatically.
You can change the high definition "Bandwidth" in the "Acquire" > "HD" dialog.
Acquire settings